Suitable for safe treatment of microbiological cultures. The INCUCELL incubator range produces no noise and provides a very soft air convection within the chamber , the variant INCUCELL V (incubator with a circulating fan) has an advantage of more precise temperature distribution with small deviations. These incubators can be used especially in biological and microbiological laboratories, quality tests in pharmacy, cosmetics and testing in veterinary medicine and food processing industry.
Volume: 22, 55, 111, 222, 404, 707 litres
Working temperature: INCUCELL: from 5°C above ambient temperature up to 100°C
INCUCELL V: from 10°C above ambient temperature up to 100°C
Inner glass door Internal Chamber: Stainless steel DIN 1.4302 (AISI 304)
The ECO Line:
Intuitive control
Microprocessor process control Fuzzy logic
Multi-lingual communication
Acoustic and visual alarm
LED indicator of device functionality
LCD display - 3 inches (7,6 cm)
Transflective brilliant FSTN display, using COG technology (it is backlit and it uses external lighting reflection – higher intensity of external light increases the display readability)
Adjustable display contrast depending on device placement
Exceptionally wide vision angle
Large signs on the display visible from afar
Current values (eg. temperature, humidity for CLIMACELL®, pressure for VACUCELL®) during the device operation are enlarged for easy readibility
Resistant foil keyboard with SoftTouch surface (pleasant to touch)
Mechanic response of keys
Lit symbols integrated directly in the foil keyboard
Keyboard lock to block unauthorised access – adjustable by multiple pressing
Real time programming and cycling (ramps as optional equipment)
Up to 9 programs, 2 segments for each program and up to 99 cycles.
USB Host port for flash disc connection for easy export of the relevant data (optional equipment)
External dimensions (including door, handle, legs N or casters K)
Width (Max)
mm
406
620
760
760
760
1160
Depth (Max)
mm
560
640
640
790
790
790
Height (Max)
mm
610N
680N
860N
1095N
1910K
1910K
ventilation neck diameter - internal / external
mm
52/49
52/49
52/49
52/49
52/49
52/49
Package - Basic
Width
Approx. mm
500
700
830
860
830
1230
Depth
Approx. mm
720
730
730
860
860
860
Height (Including Pallet)
Approx. mm
810
875
1060
1260
2085
2085
Package - Case
Width
Approx. mm
720
780
810
920
910
1335
Depth
Approx. mm
780
800
910
960
970
1060
Height (Including Pallet)
Approx. mm
835
900
1085
1310
2123
2163
Package - Wooden Crate
Width
Approx. mm
-
780
900
900
910
1310
Depth
Approx. mm
-
800
800
940
940
940
Height (Including Pallet)
Approx. mm
-
870
1090
1270
2120
2120
Trays / Shelves
Maximum Number
Pcs
4
4
7
10
19
19
Standard Number
Pcs
2
2
2
2
2
2
Minimal distance between trys/shelves
mm
60
70
70
70
70
70
Usable area
mm
185x265
380x335
520x335
520x485
520x485
920x485
Maximum allowed loading of trays*
Per 1 Tray
Kg
10
20
20
30
30
50
Per 1 Shelf
Kg
10
20
20
30
30
20
Inside the device - In Total
Kg
25
50
50
70
100
130
Number of External Metal Doors
Pcs
1
1
1
1
1
2
Weight
Net
Approx. Kg
31
55
75
100
150
215
Brut (cartoon)
Approx. Kg
36
66
87
116
175
240
Electrical Data - Mains 50/60 Hz
Max Input
kW
0.24/0.96
0.3/0.7
0.3/0.7
0.5/0.7
0.91/1.3
0.9/1.3
Stand by Input
W
5
5
5
5
5
5
Current for Voltage **
A
1.1/4.2
1.3/3
1.3/3
02-Mar
3.9/5.6
3.9/5.6
Current for Voltage **
V
230
230
230
230
230
230
IP Code
IP20
IP20
IP20
IP20
IP20
IP20
Temperature Data
Operation Temperature
From 5°C /10°C (***) above ambient temperature
to °C
100
100
100
100
100
100
Variations from operation temperature with closed flap and door (DIN 12 880 Part 2)
Space
± % Temperature
1.1/0.4
0.5/0.3
0.5/0.3
1/0.3
1/0.8
1.5/1.5
± at 37°C
0.4/0.2
0.3/0.2
0.3/0.2
0.3/0.3
0.4/0.3
0.6/0.3
Time
± °C
0.4/0.1
0.2
0.2
0.2
0.2
0.2
Time to reach temperature of 37°C with closed flap and voltage (230V)
min
70/80
49/41
57/51
79/66
41/38
59/51
Number of Air Exchanges at 37°C
per Hour
4/45
5/45
5/49
5/24
5/18
5/12
Heat Losses
at 37°C
W
20/70
30/80
45/95
45/95
65/115
85/135
Noise Level of Complete Device
dB
0
0
0
0
0
0
Note: INCUCELL®/INCUCELL® V
All the technical data apply to 22°C of ambient temperature and ± 10%.
Standard version up to 70°C, optional version up to 100°C.
*) The trays may be covered to approximately 50% of their surface and if possibly in such a way so as the air may evenly flow inside the chamber space.
**) Mains voltage is specified on type label of the device.
***) Without fan/ with fan.
The values may differ depending on specific charge and media parameters.
Need further assistance or can't find what you're looking for? Please complete all details of your enquiry and we will get back to you shortly.